Programming Guide 
11.4.3 Post Processing 
The following parameters can be set to determine the post processing 
mode, see section 6.4 
 on page 38. 
P=5 
peak hold, hold time: 5 s 
F=12.5 
valley hold, hold time: 12.5 s 
G=10 
averaging, average time (90%): 10 s 
XY=3 
advanced peak hold, hysteresis: 3 K 
XY=‐2 
advanced valley hold, hysteresis: 2 K 
Advanced Peak/Valley Hold with Averaging: 
C=250 
threshold: 250°C 
AA=15 
averaging time (90%): 15 s 
11.5 Dynamic Data 
All temperature related information is calculated 128 times a second. 
To request the dynamic data, following commands are available: 
?T 
target temperature 
?I 
internal temperature of the sensing head 
?XJ 
internal temperature of electronics housing 
?Q 
energy value of the infra‐red temperature 
?XT 
trigger set point (active/inactive) for the FTC3 input 
To  check  for  resets  (e.g.  power  shut  down)  use  the  command  XI. 
Notice, after a reset the unit is new initialized. 
?XI 
asks for the reset status 
!XI0 
no reset occurred 
!XI1 
a reset occurred, new initialization of the unit 
XI=0 
sets the reset status back to 0 
MID 