 Reference, Installation, and Operations Manual
Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
Set calibration parameters
 163
5.7.3
Set calibration parameters
The Daniel Mark III Ultrasonic Gas Flow Meter meter provides for both “dry” calibration and 
“wet” calibration. Dry calibration refers to calibration methods that use factory-set calibration 
parameters. Wet calibration refers to calibration methods that use calibration parameters from 
a flow calibration.  
1.
Select the 
Calibration|Meter Factors
 menu and Daniel MeterLink displays the Meter 
Factors dialog.
Figure 5-21  Calibration - Meter Factors Page 
Security switch S-2, Position 4 (on the CPU Board) must be closed or pushed towards the 
board before writing calibration factors to the meter.