Emerson Process Management 3422 User Manual

Page of 398
 164
Set calibration parameters
Section 5: Startup and meter configuration
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
2.
Choose a Flow Calibration Method: 
Polynomial
Piece-wise linear
None
3.
Enter the appropriate forward and reverse meter factors and click the 
Write
 button to 
send the factors to the meter.
4.
Click the 
Cancel
 button to close the dialog box. If you made changes, click the 
Write
 
button to apply the changes.  
After the meter is configured, set S-2 Position 4 (on the CPU Board) back to the open 
position to write-protect the configuration.