Emerson Process Management 3422 User Manual

Page of 398
 Reference, Installation, and Operations Manual
Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
Set calibration parameters
 165
Dry calibration
The meter uses a third-order polynomial for dry calibration as shown in 
. Click the 
Advanced View
 button to display the zero flow calibration polynomial coefficients. Regardless of 
what flow calibration method is selected, the zero flow calibration polynomial coefficients will 
be applied first.     
The meter provides separate dry calibration coefficients for each flow direction as listed in the 
table below.
Table 5-22  Data Points for Dry Calibration
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
Forward A0
Data points affected:
FwdA0
Options:
Enter a value (m/s or ft/s) within the range [-1,1 m/s].
Guidelines:
This value should only be modified under direction of Daniel personnel.
For additional guidelines, see 
 and Daniel MeterLink
 
online help.
Forward A1
Data points affected:
FwdA1
Options:
Enter a value (dimensionless) within the range [0.95,1.05].
Guidelines:
This value should only be modified under direction of Daniel personnel.
For additional guidelines, see 
 and Daniel MeterLink online help.
Forward A2
Data points affected:
FwdA2
Options:
Enter a value (s/m or s/ft) within the range [-0.1,0.1 s/m].
Guidelines:
This value should only be modified under direction of Daniel personnel.
For additional guidelines, see 
 and Daniel MeterLink online help.
Modification of dry calibration parameters is not recommended. The dry calibration 
parameters should not be modified unless directed by Daniel personnel.