SECTION 5—MAINTENANCE
34
2. Coil the air tubing.
3. Wrap the AC cord in a coil.
4. Zip the bag closed.
5. Store the device in a safe area.
NOTE: Store the repackaged device in a dry area. DO NOT place 
other objects on top of the repackaged device.
Filter Care
The accessible filter MUST be replaced after one month of use 
or when it turns GREY. The filter should be replaced if the 
nebulizer has been in storage for three months or longer.
CAUTION
Use only the filters provided with the Invacare Stratos 
Portable Nebulizer device. 
DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate without a air filter.