SECTION 6—TROUBLESHOOTING
36
Nebulizer does not work, 
and the light indicator is red.
The thermal protector has been activated due 
to:
a. The device operating outside the conditions 
outlined in this manual.
b. The device has been operating near heat 
sources or in hot environments.
Allow the nebulizer to cool down for a mini-
mum of 40 minutes. Wait until the RED light 
turns off before turning the nebulizer on again.
Optional battery is not 
working optimally.
Battery was not properly conditioned (charged 
for at least 16 hours before first use).
If it has been in use for approximately four 
months, it may be time to recondition the bat-
tery.
Optional battery is not 
charging.
Battery will not charge if connected to the nebu-
lizer.
Check all power cord connections and cord 
conditions.
Check power outlet.
PROBLEM
SOLUTION