syngo Dynamic Evaluation
 
317
General Hints
• Slice Position with Multi-Row Scanners
When you use a multi-row scanner, several slice posi-
tions are scanned per acquisition. When you load the 
images, the window Slice Selection is displayed, in 
which you can define the slice position to be analyzed. 
You can evaluate either the scanned examination data 
or an averaged slice position.
• Help to Interpret the Enhancement
Parameter images that help to interpret the enhance-
ment are displayed in segment 2, for example, the MIP 
image and the Peak Enhancement image.
• Defining the Tissue Range
With Tissue Classification you can exclude all tissue 
regions outside the CT-value range entered. This 
allows you to emphasize the organ to be examined and 
remove disturbing factors. 
HU limitation does not affect evaluation of the ROIs 
and only appears at the Time to Peak and the Peak 
Enhancement image.
The excluded tissue densities are masked by a color in 
the baseline image. The excluded areas are displayed 
in black in the Time to Peak and Peak Enhancement 
images.