5 MEASUREMENT 
This meter is capable of taking measurements with automatic or manual temperature 
compensation. Automatic temperature compensation only occurs when a 
temperature sensor is plugged into the meter. If there is no temperature sensor 
plugged into the meter, the default manual temperature setting is automatically 25.0 
°C.  You can manually set the temperature to match your working conditions using a 
separate thermometer. 
NOTE: Remove the protective rubber cap or soaker bottle of the electrode before 
proceeding with measurement. Take care not to exert too much force as this may 
cause damage to the electrode. 
5.1 Automatic 
Temperature 
Compensation 
For automatic temperature compensation (ATC) simply 
plug the temperature probe into the meter. The ATC 
indicator will light up on the LCD.  
 
NOTE:  If you are using a temperature probe, the probe 
must be submersed in the liquid you are measuring.  
 
 
MEAS
ATC
°C
pH
7.05
24.5
Figure 15: ATC indicator