CyberScan pH 11 / 110  
 
19
5.3 Taking 
Measurements 
Be sure to remove the electrode soaker bottle or protective rubber cap on the 
electrode before measurement.  
To take readings: 
1.  Rinse the probe with de-ionised or distilled 
water before use to remove any impurities 
adhering to the probe body. If the pH electrode 
has dehydrated, soak it for 30 minutes in 
electrode storage solution or 2M – 4M KCl 
solution (sold separately). 
2.  Press ON to switch on meter. The MEAS 
annunciator appears on the top center of the 
LCD. The ATC indicator appears in the lower 
right-hand corner to indicator Automatic Temperature Compensation (section 
5.2 on Manual Temperature Compensation). 
3.  Dip the probe into the sample. 
When dipping the probe into the sample, the sensor or the glass bulb  of  the         
electrode must be completely immersed into the sample. Stir the probe gently in the 
sample to create a homogeneous sample. 
4.  Allow time for the reading to stabilise. Note the reading on the display. 
5.  To toggle between pH and mV (or Rel mV) readings, press the MODE key. 
5.3.1  Taking measurements with READY indicator selected on 
If the READY indicator has been activated, the READY annunciator lights when the 
reading is stable
2
 *. You can switch the READY indicator on or off in SETUP 
program. See 9.2.1 - P1.1: READY Indicator and Auto Hold function on page 34.  
5.3.2  Taking measurements with the AUTO HOLD feature selected on 
This feature is available on model pH 110 only.  
When a reading is stable for more than 5 seconds, the AUTO HOLD feature will 
automatically “HOLD” the reading. The “HOLD” indicator appears on the left side of 
the display. Press the HOLD/ENTER key to release the reading.  Switch the Auto 
Hold feature on or off in SETUP program in section 9.2.1 - P1.1: READY Indicator 
and Auto Hold function on page 34. 
                                                           
2
 The READY indicator appears and the reading holds until the measured value exceeds the tolerance (±0.02 pH; 
±0.2 mV <400; ±2 mV > 400). Then, the READY annunciator turns off. 
MEAS
pH
ATC
°C
8.23
21.3
Figure 17: Measurement 
mode