eutech-instruments cyberscan ph 110 phmvrs232 meter User Manual

Page of 76
Instruction Manual   
 
CyberScan pH 11 / 110  
 
18
5.2 
Manual Temperature Compensation 
IMPORTANT: For manual compensation, you must disconnect the temperature 
probe. The [ATC] annunciator will disappear from the LCD.  
1.  Switch the meter on. Press the MODE key to 
select pH mode.  
2.  Press the CAL/MEAS key to enter pH 
calibration mode. The CAL indicator will appear 
above the primary display.  
3.  While in pH calibration mode, press the MODE 
key to enter temperature calibration mode. The 
primary display act as the adjustable 
temperature setting and the secondary display 
shows the default temperature value of 25°C or 
indicates the last set temperature setting. 
4.  Check the temperature of your sample using an 
accurate thermometer.  
5.  Press the MI/▲ or MR/▼ keys to set the 
temperature to the measured value from step 4.  
6.  Press HOLD/ENTER key to confirm the 
selected temperature and to return to the pH 
measurement mode.  
7.  The meter will now compensate pH readings for 
the manually set temperature. 
See Figure on right.  
 
NOTES: 
To exit this program without confirming the manual temperature 
compensation value, DO NOT press ENTER in step 6.  Press 
CAL/MEAS instead. 
 
 
  
MEAS
pH
°C
7.05
30.0
CAL
°C
25.0
25.0
°C
30.0
25.0
CAL
READY
MI
MR
ENTER
HOLD
Figure 16: Manual 
temperature compensation