eutech-instruments carbon dioxide gas User Manual

Page of 20
Instruction Manual 
 
Carbon Dioxide Electrode
 
 
17
TROUBLESHOOTING HINTS 
 
Symptom                     
Possible Causes          
 
Next Step 
Out of Range    
 
defective meter           
 
perform meter checkout 
Reading 
 
       procedure 
(see 
meter 
 
 
 
 
 
 
 
 
 
instruction manual) 
 
 
 
 
   defective 
inner 
body 
  refer 
to 
Checking the 
                                             
 
 
 
 
Electrode Inner Body
 
 
                          
 
electrode not           
 
 
unplug electrode and reset 
                          
 
plugged in properly   
 
 
                          
 
internal filling        
 
 
fill outer body  of electrode 
                          
 
solution not added        
 
with proper amount of  
                                                 
 
 
 
 
internal filling solution 
 
                          
 
air bubble on membrane          
 
remove bubble by 
                         
 
               
 
 
 
re-dipping electrode 
                                               
                          
 
electrode not in solution 
 
put electrode in solution 
 
                    
Noisy or Unstable         
insufficient internal          
 
fill outer body of electrode 
Reading (readings         
filling solution      
 
 
with proper amount of  
continuously or                      
 
 
 
 
internal filling solution 
random changing.)                               
                          
 
defective meter           
 
perform meter checkout 
                                                 
 
 
 
 
procedure (see meter 
                                                 
 
 
 
 
instruction manual) 
 
                          
 
bottom cap loose          
 
ensure that bottom cap is 
                                                 
 
 
 
 
screwed on tight enough 
                                                 
 
 
 
 
to close gap between 
                                                 
 
 
 
 
bottom cap and body 
 
                          
 
defective inner body     
 
refer to 
Checking the 
                         
 
 
 
 
 
 
Electrode Inner Body 
 
                          
 
air bubbles on membrane         
remove bubbles by 
                                
 
 
 
 
 
 
redipping electrode 
 
                          
 
meter or stirrer        
 
 
check meter and stirrer 
                          
 
improperly grounded    
 
for grounding 
 
 
Drift (reading slowly     
internal filling solution  
 
ensure that membrane is  
changing in one            
leakage 
 
 
 
installed properly 
direction)