eutech-instruments carbon dioxide gas User Manual

Page of 20
Instruction Manual 
 
Carbon Dioxide Electrode
 
 
18
                         
 
incorrect internal filling     
 
refill outer body of electrode 
                           
 
solution          
 
 
using filling solution shipped 
                                                 
 
 
 
 
with electrode 
 
                          
 
total sample level       
 
dilute sample of dissolved 
                         
 
 
 
 
 
 
species above 1M 
 
                          
 
electrode in sample      
 
reduce surface area-to- 
                          
 
too long; CO
2
 loss        
 
volume ratio, slow down rate  
                                                  
 
 
 
 
of stirring, avoid high 
                                                 
 
 
 
 
temperatures 
 
                          
 
membrane failure         
 
replace membrane 
                          
 
(wet, perforation, discoloration) 
 
                          
 
samples and standards              
allow samples and standards 
                          
 
not at constant       
 
 
to come to room temperature 
                                        
temperature 
 
 
          before use 
 
                          
 
heat generated by         
 
place insulating material 
                          
 
magnetic stirrer       
 
 
between stirrer and beaker 
 
                          
 
defective inner body        
 
refer to 
Checking the 
                          
 
 
 
 
 
 
Electrode Inner Body 
 
                           
 
electrode exposed to     
 
hold electrode by outer body 
                           
 
air for extended period    
 
and pull on electrode cable. 
               
 
 
 
 
 
 
 
Internal filling solution will 
                                                 
 
 
 
 
flow under membrane and 
                                                
 
 
 
 
restore electrode response 
 
Low Slope or No Slope      
standards contaminated              
prepare fresh standards  
                           
 
or incorrectly made 
                           
 
 
    buffer 
not 
used   use 
recommended 
buffer 
 
                           
 
standard used as         
 
use buffer 
                           
 
buffer 
 
                           
 
electrode exposed        
 
hold electrode by outer body 
                            
 
to air for extended       
 
and pull on electrode cable. 
                           
 
period                                       
Internal filling solution will 
                                                 
 
 
 
 
flow under membrane and 
                                                                                    
 
restore electrode response 
                                                 
                          
 
membrane failure         
 
replace membrane 
                           
 
(wet, perforation, discoloration) 
                            
                          
 
defective inner body      
 
refer to 
Checking the 
                       
 
 
 
 
 
 
Electrode Inner Body