148 
DR4300 Circular Chart Recorder Product Manual 
12/03 
5.4.3  Thermocouple Inputs Using an Ice Bath or Ice Point Reference 
 
Table 5-6  Set Up Wiring Procedure for Thermocouple Inputs Using an Ice Bath 
Step Action 
Connect the copper leads to the calibration source (see Figure 5-3). 
Connect a length of thermocouple extension wire to the end of each copper lead and insert the 
junction points into the ice bath. 
Connect the other end of the thermocouple extension wires to the TB2 terminals on the printed 
circuit assembly for the input channel being calibrated.  
 
Main PCA 
Pen 2
Main PCA 
Pen 1
7
TB2
TB1
7
TB2
TB1
TB2
+
R
-
+
mV source
Copper
wires
Thermocouple
extension wires
Containers of 
crushed ice
24092
 
Figure 5-3  Calibration Set Up Diagram for Thermocouple Inputs Using an Ice Bath