Language Dictionary - 6
57
 
Calibration Commands
 
Calibration commands let you enable and disable the calibration mode, change the calibration password,
calibrate current and voltage programming, and store new calibration constants in nonvolatile memory.
Only one output channel may be calibrated at a time.
NOTE:
If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error. You must also save any changes that you
made using CALibrate:SAVE, otherwise all changes will be lost when you exit
calibration mode.
 
CALibrate:CURRent
 
This command initiates the calibration of the current priority mode as well as the 0.5mA current range
measurement circuit.
   
 
Command Syntax
 
CALibrate:CURRent[:LEVel] (@<channel>)
 
Parameters
 
None
 
Examples
 
CAL:CURR(@1)
! start current calibration
 
Related Commands
 
CAL:CURR:LIM,
CAL:CURR:MEAS
 
 
CALibrate:CURRent:LIMit[:POSitive]
CALibrate:CURRent:LIMit:NEGative
 
This command initiates the calibration of the positive or negative current limit.
   
 
Command Syntax
 
CALibrate:CURRent:LIMit[:POSitive] (@<channel>)
CALibrate:CURRent:LIMit:NEGative (@<channel>)
 
Parameters
 
None
 
Examples
 
CAL:CURR:LIM (@1)
CAL:CURR:LIM:NEG (@1)
 
Related Commands
 
CAL:CURR
CAL:CURR:MEAS
 
 
CALibrate:CURRent:MEASure
 
This command initiates the calibration of the 0.5A or 15mA current range measurement circuit.
   
 
Command Syntax
 
CALibrate:CURRent:MEASure <NRf>, (@<channel>)
 
Parameters
 
A value that falls within the 0.5A or 15mA current range
 
Unit
 
(amperes)
 
Examples
 
CAL:CURR:MEAS 0.5,(@1)
!0.5A range
CAL:CURR:MEAS 0.005,(@1)
!15mA range
 
Related Commands
 
CAL:CURR,
CAL:CURR:LIM