Language Dictionary - 6
59
 
CALibrate:SAVE
 
This command saves any new calibration constants after a calibration procedure has been completed in
nonvolatile memory. If CALibrate:STATe OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored..
   
 
Command Syntax
 
CALibrate:SAVE
 
Parameters
 
None
 
Examples
 
CAL:SAVE
 
Related Commands
 
CAL:PASS      CAL:STAT
 
 
CALibrate:STATe
 
This command enables and disables calibration mode. The calibration mode must be enabled before the
dc source will accept any other calibration commands.
 
The first parameter specifies the enabled or disabled state. The second parameter is the password. A
password is required if calibration mode is being enabled and the existing password is not 0.  If the
password is not entered or is incorrect, an error is generated and the calibration mode remains disabled.
The query returns only the state, not the password.
NOTE:
Whenever the calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
   
 
Command Syntax
 
CALibrate:STATe <bool>[,<NRf>]
 
Parameters
 
0 | OFF | 1 | ON [,<password>]
 
*RST Value
 
OFF
 
Examples
 
CAL:STAT 1,3280
CAL:STAT OFF
 
Query Syntax
 
CALibrate:STATe?
 
Returned Parameters
 
<NR1>
 
Related Commands
 
CAL:PASS     CAL:SAVE     *RST
 
 
CALibrate:VOLTage
 
This command initiates the calibration of the output voltage and the voltage measurement circuit.
   
 
Command Syntax
 
CALibrate:VOLTage (@<channel>)
 
Parameters
 
None
 
Examples
 
CAL:VOLT (@1)