6 – Language Dictionary
58
 
CALibrate:DATA
 
This command enters a calibration value that you obtain by reading an external meter. You must first
select a calibration level (with CALibrate:LEVel) for the value being entered.
   
 
Command Syntax
 
CALibrate:DATA <NRf>
 
Parameters
 
<external reading>
 
Unit
 
or (amperes or volts)
 
Examples
 
CAL:DATA 3222.3 MA
CAL:DATA 5.000
 
Related Commands
 
CAL:STAT     CAL:LEV
 
 
CALibrate:DATE
 
This command stores the date the unit was last calibrated. Enter any ASCII string up to 31 characters.
   
 
Command Syntax
 
CALibrate:DATE <date>
 
Parameters
 
<date>
 
Examples
 
CAL:DATE "3/22/01"
CAL:DATE "22.3.99"
 
Query Syntax
 
CALibrate:DATE?
 
Returned Parameters
 
<SRD>
 
 
CALibrate:LEVel
 
This command selects the next point in the calibration sequence. P1 is the first calibration point,
P2 is the second calibration point.
   
 
Command Syntax
 
CALibrate:LEVel <point>
 
Parameters
 
P1 | P2
 
Examples
 
CAL:LEV P2
 
 
CALibrate:PASSword
 
This command lets you change the calibration password. A new password is automatically stored in
nonvolatile memory and does not have to be stored with CALibrate:SAVE. The default password is the
model number of the unit. If the password is set to 0, password protection is removed and the ability to
enter the calibration mode is unrestricted.
   
 
Command Syntax
 
CALibrate:PASSword <NRf>
 
Parameters
 
<model number> (default)
 
Examples
 
CAL:PASS 1234
 
Related Commands
 
CAL:SAV