Precision Trading PM5200 Manuale Utente

Pagina di 16
11
Blender
Air-Oxygen
3. To activate the bleed, turn the knurled collar until the bleed 
pin is engaged. Then slide the collar back until it contacts the 
cover.
4. Confirm the concentration of air/oxygen with an Oxygen 
Analyzer/monitor.
5. Confirm the flow of air and/or oxygen mixture to the patient.
6. Turn “OFF” the Air and Oxygen supply when deactivating the 
Auxiliary bleed, and no device is attached to the outlet port.
7. To deactivate the bleed, pull the collar out away from the cover 
while rotating the collar until the slot engages the pin. Then 
slide the collar past the pin and turn an additional 
1
/
4
 turn to 
lock in place.
NOTE: This Procedure is completed with ease when all gas 
supplies to the Blender are turned “OFF”.
8. Turn “OFF” the Air and Oxygen supply or disconnect when 
the Blender is not in use.
9. Operational Verification listed below should be done before 
the Blender is placed in service.
Verification consists of:
• Alarm Test
• Verify FiO
Concentration with
 
a Oxygen Analyzer/monitor
• Reverse Gas Flow Procedure
PERFORMANCE CHECk
A detailed description of the tests listed below can be found in the 
Blender Service manual (P/N 504827), available on the Internet; 
www.precisionmedical.com
• Reverse Gas Flow Procedure
• Operational Verification Procedure (OVP)
CLEANiNg
CAUTiON
dO NOT steam autoclave.
dO NOT immerse the Air-Oxygen Blender into any liquid.
dO NOT use any strong solvent or abrasive cleaners.
dO NOT gas sterilize with (ETO) Ethylene Trioxide.
dO NOT clean with aromatic hydrocarbons.