Blender
Air-Oxygen
TROUBLESHOOTiNg
If the Air-Oxygen Blender fails to function, consult the 
Troubleshooting Guide below.
If  problem  cannot  be  solved  by  using  Troubleshooting  Guide 
refer  to  the  Air-Oxygen  Blender  Service  manual  (P/N  504827) 
available on the Internet; www.precisionmedical.com or consult 
your Provider.
Problem
Probable Cause
Remedy
Oxygen 
concentration 
discrepancy 
between Blender 
setting and 
Analyzer/monitor 
(greater than 3%)
1. High Flow model, 
flow requirement 
below 15 lpm.
Low Flow model,
flow requirement 
below 3 lpm.
2. Analyzer/monitor 
inaccurate
3. Low flow bleed 
obstructed
4. Gas supply 
contaminated
5. Downstream device 
causing back flow 
or restricted flow
1. Use auxiliary outlet 
& engage bleed
2. Recalibrate 
Analyzer/monitor or 
Verify with second 
Analyzer/monitor
3. Remove obstruction
4. Check gas sources 
with calibrated 
Oxygen Analyzer/
monitor to confirm 
Oxygen is 100% and 
Air is 21%
5. Isolate Blender. 
Check oxygen 
concentration at 
Blender Outlets
No flow at
Blender outlets
1. Gas sources turned 
“OFF”
2. Gas sources not 
connected
1. Turn gas sources 
“ON”
2. Connect gas 
sources
 Alarm sounding
1. Difference between 
Oxygen and air inlet 
pressures greater 
than specified
1. Correct pressure 
difference until 
Air and Oxygen 
pressures are within 
specification