Precision Trading PM5200 Manuale Utente

Pagina di 16
12
Blender
Air-Oxygen
1. Disconnect all gas connections and equipment before 
cleaning. 
2. Clean exterior surfaces with a cloth dampened with mild 
detergent and water.
3. Wipe dry with a clean cloth.
MAiNTENANCE
The following maintenance on the Air-Oxygen Blender must be 
performed by a trained service technician:
Every month verify proper functioning of alarm.
Every year conduct the Operational Verification Procedure (OVP).
Every 2 years the Air-Oxygen Blender should be serviced. 
PM5200 (P/N 505407) 
PM5300 (P/N 504932)
Refer to the Air-Oxygen Blender Service manual (P/N 504827) 
for  complete  details  regarding  further  maintenance  and 
testing.
TECHNiCAL dESCRiPTiON
For a complete Technical Description of the Air-Oxygen Blender 
and list of Replacement Parts, reference the Air-Oxygen Blender 
Service manual (P/N 504827) available on the Internet; 
www.precisionmedical.com.
RETURNS
Returned  products  require  a  Returned  Goods  Authorization 
(RGA) number, contact Precision medical, Inc. All returns must 
be packaged in sealed containers to prevent damage. Precision 
medical, Inc. will not be responsible for goods damaged in transit. 
Refer  to  Precision  medical,  Inc.  Return  Policy  available  on  the 
Internet; www.precisionmedical.com.
diSPOSAL iNSTRUCTiONS
This device and its packaging contain no hazardous materials. 
No special precautions need to be taken when disposing the 
device and/or its packaging.
Please Recycle