Agilent Technologies E6380-90073 Manuale Utente
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
Chapter 1
153
GPIB Command Dictionary
CONFigure subsystem
:NOTChmode <string>
:NOTChmode?
:NOTChmode?
These commands set/query the coupling between the AF Generator 1
and a notch filter. The frequency and gain of the notch filter is set by
the AFAN:NOTC:GAIN and AFAN:NOTC:FREQ commands.
and a notch filter. The frequency and gain of the notch filter is set by
the AFAN:NOTC:GAIN and AFAN:NOTC:FREQ commands.
Syntax
CONF:NOTC ‘AFGen1’ !notch filter switched in
‘None’
!no notch filter
CONF:NOTC?
Screen/field equivalent
CONF:NOTC controls the
Notch Coup l
field of the INSTRUMENT
CONFIGURE screen.
:OFLevel:MODE <string>
:OFLevel:MODE?
:OFLevel:MODE?
These commands set/query the RF level offset, and is typically used to
compensate for path loss between the Test Set and the base station.
compensate for path loss between the Test Set and the base station.
Syntax
CONF:OFL:MODE ‘On’
‘Off’
CONF:OFL:MODE?
Screen/field equivalent
CONF:OFL:MODE controls the
RF Level Offset
field of the