Chapter 1
259
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer:POWer[:VALue]?
IS-2000 only. This query returns an array of 128 values that represent 
the channel power relative to the selected reference channel power in a 
Power & Noise measurement.
NOTE
The values returned represent the relative power of individual Walsh 
channels not the supplemental channels of which they may be a part. 
The power of supplemental channels in a Power & Noise measurement 
must be read directly from the 
Marker Lvl
 value on the Code Domain 
Analyzer screen.
Syntax
MEAS:CDAN:IS2000:TRAC:POW:POW[:VAL]?
Screen/field equivalent
MEAS:CDAN:IS2000:TRAC:POW:POW[:VAL]? displays information 
(top value of each bar) for the power and noise measurement when the 
CD pwr unit 
is set to 
Rel
 on the 
Reference 
menu of the CODE 
DOMAIN screen when the CDMA standard is set to “IS-2000”.
:DEC:AMPS|TACS:NBITs?
These commands set/query the number of bits measured by the 
signaling decoder in AMPS-TACS mode.
Syntax
MEAS:DEC:AMPS:NBIT?
Example
MEAS:DEC:AMPS:NBIT? !returns an integer value reporting the 
number of bits measured by the signaling unit.
For example, if the Signaling Decoder displays a value of 928 
in the 
Num of Bits 
field of the SIGNAL DECODER screen, the 
value 928 will be returned by the query.
Screen/field equivalent
MEAS:DEC:AMPS:NBIT? reads the 
Num of Bits
 measurement field 
on the SIGNALING DECODER screen when the 
Mode
 field is set to 
AMPS-TACS
.