Chapter 1
 
GPIB Command Dictionary
SANalyzer subsystem
:RFGenerator <string>
:RFGenerator?
This command selects between the fixed frequency generator and the 
tracking generator. The query form returns the mode setting of the 
generator.
Syntax
SAN:RFG ‘Track’
 ‘Fixed’
SAN:RFG?
Screen/field equivalent
SAN:RFG controls the 
Track/Fixed
 field on the 
RF Gen
 menu of the 
SPEC ANL screen.
:RLEVel <real number>
:RLEVel?
These commands set/query the reference level of the spectrum analyzer.
This command utilizes the “Real Number Setting Syntax” on page  299, 
but does not use the :STATe command.
Syntax
SAN:RLEV <real number>
SAN:RLEV?
Screen/field equivalent
SAN:RLEV controls the 
Ref Level
 field on 
Main
 menu of the SPEC 