SANalyzer subsystem
:SPAN <real number>
:SPAN?
These commands set/query the span of the spectrum analyzer display.
This command utilizes the “Real Number Setting Syntax” on page  299, 
but does not use the :STATe command.
Syntax
SAN:SPAN <real number>
SAN:SPAN?
Screen/field equivalent
SAN:SPAN controls the 
Span
 field on the 
Main
 menu SPEC ANL 
screen.
:TGENerator:AMPLitude <real number>
:TGENerator:AMPLitude?
These commands set/query the amplitude of the tracking generator. To 
select the tracking generator use the SAN:RFG ‘Track’ command.
Syntax
SAN:TGEN:AMPL <real>
SAN:TGEN:AMPL?
Screen/field equivalent
SAN:TGEN:AMPL controls the 
Amplitude
 field on the 
RF Gen
 menu of 
the SPEC ANL screen. The 
Track/Fixed
 field must be set to 
Track
.