Chapter 1
GPIB Command Dictionary
SANalyzer subsystem
:TGENerator:DESTination <string>
:TGENerator:DESTination?
:TGENerator:DESTination?
These commands set/query the output port for the tracking generator.
To select the tracking generator use the SAN:RFG ‘Track’ command.
To select the tracking generator use the SAN:RFG ‘Track’ command.
Syntax
SAN:TGEN:DEST ‘RF Out’
‘Dupl’
SAN:TGEN:DEST?
Screen/field equivalent
SAN:TGEN:DEST controls the RF Out/Dupl subfield of the
Port/Sweep
field on the
RF Gen
menu of the SPEC ANL screen. The
Track/Fixed
field must be set to
Track
.
:TGENerator:OFRequency <real number>
:TGENerator:OFRequency?
:TGENerator:OFRequency?
These commands set/query the offset frequency of the tracking
generator. To select the tracking generator use the SAN:RFG ‘Track’
command.
generator. To select the tracking generator use the SAN:RFG ‘Track’
command.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
but does not use the :STATe command.
Syntax
SAN:TGEN:OFR <real number>
SAN:TGEN:OFR?
Screen/field equivalent
SAN:TGEN:OFR controls the
Offset Freq
field on the
RF Gen
menu of
the SPEC ANL screen. The
Track/Fixed
field must be set to
Track
.