GPIB Command Dictionary
CANalyzer subsystem
CANalyzer subsystem
The CANalyzer subsystem controls functions unique to the CDMA 
ANALYZER screen.
:ACP:POWer:CALibrate
This command initiates calibration of the adjacent channel power 
measurement for the current RF analyzer frequency at the selected 
frequency offset and filter bandwidth.
Syntax
CAN:ACP:POW:CAL
Example
CAN:ACP:POW:CAL 
!calibrates ACP measurement
Screen/field equivalent
CAN:ACP:POW CAL controls the 
ACP Cal
 field on the CDMA 
ANALYZER screen.
:ACP:POWer:FILTer <real number>
:ACP:POWer:FILTer?
This command selects the filter bandwidth over which the adjacent 
channel power measurement is made in the adjacent channels. Valid 
values are 10 kHz to 1.23 MHz with 500 Hz resolution. 
Syntax
CAN:ACP:POW:FILT <real number>
CAN:ACP:POW:FILT?
Example
CAN:ACP:POW:FILT 10 kHz
!sets the filter BW to 10 kHz
CAN:ACP:POW:FILT?
!returns the filter BW setting
Screen/field equivalent
CAN:ACP:POW:FILT controls the 
ACP Fltr BW
 field on the CDMA 
ANALYZER screen.