5-10
 
Optimizing Measurement Results
Increasing Sweep Speed
Increasing Sweep Speed
You can increase the analyzer sweep speed by avoiding the use of some features that 
require computational time for implementation and updating, such as bandwidth marker 
tracking.
You can also increase the sweep speed by making adjustments to the measurement 
settings. The following suggestions for increasing sweep speed are general rules that you 
should experiment with:
• use swept list mode
• decrease the frequency span
• set the auto sweep time mode
• widen the system bandwidth
• reduce the averaging factor
• reduce the number of measurement points
• set the sweep type
• use chop sweep mode
• use external calibration
• use fast 2-port calibration mode
To Use Swept List Mode 
When using a list frequency sweep, choosing swept list mode can increase throughput by 
up to 6 times over stepped list mode. This mode takes data while sweeping through each 
list segment. In addition, this mode expands the list table to include test port power and IF 
bandwidth. Selectable IF bandwidths can increase the throughput of the measurement by 
allowing the user to specify narrow bandwidths only where needed.
• For in-depth information on swept list mode, refer to 
.
• For more information on making measurements with swept list mode, refer to 
.
1. To set up a swept list measurement, press 
 
 
 
.
2. The frequency segments can be defined in any of the following terms:
• start/stop/number of points/power/IFBW
• start/stop/step/power/IFBW
• center/span/number of points/power/IFBW
• center/span/step/power/IFBW
3. When finished, press 
 
.
Sweep Setup
 SWEEP TYPE MENU 
 EDIT LIST   ADD 
 DONE   LIST TYPE: [SWEPT]