Operating Concepts
Sweep Types
The frequency subsweeps, or segments, can be defined in any of the following terms:
• start/stop/number of points
• start/stop/step
• center/span/number of points
• center/span/step
• CW frequency
The subsweeps can overlap, and do not have to be entered in any particular order. The 
analyzer sorts the segments automatically and lists them on the display in order of 
increasing start frequency, even if they are entered in center/span format. If duplicate 
frequencies exist, the analyzer makes multiple measurements on identical points to 
maintain the specified number of points for each subsweep. The data is shown on the 
display as a single trace that is a composite of all data taken. The trace may appear uneven 
because of the distribution of the data points, but the frequency scale is linear across the 
total range.
Once the list frequencies have been defined or modified, the list frequency sweep mode can 
be selected with the 
 softkey in the sweep type menu. The 
frequency list parameters can also be saved with an instrument state.
Swept List Frequency Sweep (Hz) 
The 
 softkey activates a swept list frequency sweep, one of two list 
frequency sweep modes. The swept list mode allows the analyzer to sweep a list of 
arbitrary frequency points which are defined and modified in a way similar to the stepped 
list mode.
However, this mode takes data while sweeping through the defined frequency points, 
increasing throughput by up to 6 times over a stepped sweep. In addition, this mode allows 
the test port power and IF bandwidth to be set independently for each segment that is 
defined. The only restriction is that you cannot specify overlapping frequency segments.
Similar to stepped list mode, the 
 softkey also provides access to the 
segment menu. However, swept list mode expands the way segments can be defined.Refer 
to the following information on how to enter or modify the list segments.
Swept Edit List Menu 
The 
 softkey within the sweep type menu provides access to the edit list menu. 
The function of this menu is the same as in the stepped list mode.
Swept Edit Subsweep Menu 
Using the 
 or 
 softkey within the edit list menu will display the edit subsweep 
menu. This menu lets you select measurement frequencies arbitrarily. Using this menu it 
is possible to define the exact frequencies to be measured on a point-by-point basis at 
specific power levels and IF bandwidth settings. The total sweep is defined with a list of 
subsweeps.
LIST FREQ [STEPPED] 
LIST FREQ [SWEPT] 
LIST FREQ [SWEPT] 
EDIT LIST 
EDIT 
ADD