Making Measurements
Using Display Functions
Once made active, a channel can be configured independently of the other channels in most 
variables except stimulus. For example, once channel 3 is active, you can change its format 
to a Smith chart by pressing 
 
.
Customizing the Four-Channel Display 
When one or both auxiliary channels are enabled, 
 and 
 interact to produce different display configurations according to 
.
Channel Position Softkey
 gives you options for arranging the display of the channels. Press 
 to use 
.
 works with 
. When 
 is 
selected, 
 gives you two choices for a two-graticule display:
• Channels 1 and 2 overlaid in the top graticule, and channels 3 and 4 are overlaid in the 
bottom graticule.
• Channels 1 and 3 are overlaid in the top graticule, and channels 2 and 4 are overlaid in 
the bottom graticule.
When 
 is selected, 
 gives you two choices for a 
four-graticule display:
• Channels 1 and 2 are in separate graticules in the upper half of the display, channels 3 
and 4 are in separate graticules in the lower half of the display.
• Channels 1 and 3 are in the upper half of the display, channels 2 and 4 are in the lower 
half of the display.
Table 1-1 Customizing the Display 
Split Display 
Dual Channel 
Aux Channels On 
Number of Graticules 
1X
Don’t Care
Don’t Care
1
1X/2X/4X
Off
None
2X/4X
Off
3 or 4
2
2X
On
Don’t Care
4X
On
3 or 4
3
4X
On
Both on
4
Format
 SMITH CHART 
 DUAL CHAN on OFF 
 SPLIT DISP 1X 2X 4X 
 CHANNEL POSITION 
Display
 DUAL|QUAD SETUP 
 CHANNEL POSITION 
 CHANNEL POSITION 
 SPLIT DISP 1X 2X 4X 
 SPLIT DISP 2X 
 CHANNEL POSITION 
 SPLIT DISP 4X 
 CHANNEL POSITION