Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
6- 65
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
Perform the Confidence Check
The confidence check is a means of visually checking the quality of the calibration. The 
confidence check displays the currently measured data (DATA trace) and the 
factory-premeasured data (MEM trace) for the module’s confidence state. The confidence 
state, an independent reference standard, was not used for the calibration. For a good 
calibration measurement, there should be no significant difference between the traces (see 
To Perform the Confidence Check
NOTE
A confidence check is only valid after an ECal has been performed.
1. Press 
  
.
2. Press 
 until the correct module (A or B) is selected.
3. Select 
 to display the confidence check and the ECal 
Confidence Menu.
4. Press 
 until the S-parameter that you want to view is displayed.
Pressing the 
 softkey toggles between the S11S21S12, and S22 (or 
for the ET models, TRANS (transmission) and REFL (reflection)) calibration traces. 
The confidence check will only display the S-parameter calibration data for which you 
calibrated. Uncalibrated S-parameter data traces are invalid. The memory trace still 
displays the module’s factory-premeasured trace.
For example, 
If you performed:
You can view:
A 1-port calibration
 
(
ES models)
 
(
ES models)
 (ET models)
Only one calibrated S-parameter
An enhanced response calibration
 
(
ES models)
 
(
ES models)
 (ET models)
Only two calibrated S-parameters, one 
at a time
A full 2-port calibration
 
(
ES models)
All four calibrated S-parameters, one 
at a time
Cal
 ECal MENU   CONFIGURE [  ] 
 MODULE A b 
 CONFIDENCE CHECK 
 PARAMETER [  ] 
 PARAMETER [  ] 
 S11 1-PORT 
 S22 1-PORT 
 REFLECTION 1-PORT 
 S11/S21 ENH. RESP. 
 S22/S12 ENH. RESP. 
 TRANS/REFL ENH. RESP. 
 FULL 2-PORT