Agilent Technologies 22ET 8719ES Manuale Utente

Pagina di 478
6- 67
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
Investigating the Calibration Results Using the ECal Service Menu
CAUTION
The confidence check described in the previous section displays the ECal data 
of a single state. This confidence state is a calibrated standard not used 
during ECal. It is provided to give an independent assessment of the quality 
of a calibration. In the Ecal Service menu, you may also display each of the 
calibration standards which are used during an ECal calibration, along with 
the analyzer’s measurement of those standards.
You may notice a difference in measurement results when comparing ECal 
confidence state data and ECal standard state data. This result may be 
related to certain measurement errors with the network analyzer system, 
which add uncertainty to measurement results. More detailed information 
regarding measurement uncertainty is documented in the “Determining 
System Measurement Uncertainties” chapter of the network analyzer's 
reference guide. Additional information on improving your measurements 
can be found in 
Using the ECal Service menu is not a standard part of the ECal procedure. It is a tool to 
allow you to identify problems in the calibration equipment, cables, connectors, or 
procedures.
The Confidence Check menu supports the comparison of the measured data versus the 
module's premeasured calibration data for the confidence state. The ECal Service menu 
supports the comparison of the measured data versus the module's premeasured 
calibration data for the remainder of the module states, in addition to the confidence state.
Access the ECal Service menu by pressing 
 from the Confidence Check 
menu.
The ECal Service menu softkeys are:
 
Toggles the analyzer to show the data for the following calibration 
states:
• CONF (Confidence)
• THRU 
• ISOL (Isolation)
• S11 REFL (S11 Reflection)
• S22 REFL (S22 Reflection)
 
Toggles between the available S11 and S22 reflection states listed in 
the 
 softkey described above. 
• S11 REFL allows up to 13 reflection states. 
• S22 REFL allows up to 13 reflection states. 
 ECal SERVICE
 
 
 ECal STD [  ]
 
 
 REFL STD [  ]
 
 
 ECal STD [  ]