Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
Perform the 2-Port Error Corrections
Figure 6-27 Two-Port Cal Set 1
2. Connect the ECal module between adapter A1 and adapter A3.
3. Press
.
4. Press
to perform the first 2-port error correction using the ECal
module.
NOTE
When using adapter removal calibration, you must save calibration sets to
the internal disk, not to internal memory.
the internal disk, not to internal memory.
5. Save the results to disk. Name the file "PORT1."
6. Remove the ECal module and adapter A3 from the setup. Connect adapter A3 to
adapter A1 on port 1 as shown in
Cal
ECal MENU MODULE A b
FULL 2-PORT