7- 19
Operating Concepts
Sweep Types
Setting Segment IF Bandwidth 
To enable the 
 function, you must first select 
 in 
the edit subsweep menu. List IF bandwidth is off by default and the asterisks that appear 
in the "IFBW" column of the list table indicate that the IF bandwidth for the sweep is 
being set by the normal analyzer controls.
Narrow IF bandwidths require more data samples per point and thus slow down the 
measurement time. Selectable IF bandwidths can increase the throughput of the 
measurement by allowing you to specify narrow bandwidths only where needed.
Power Sweep (dBm) 
The 
 softkey turns on a power sweep mode that is used to characterize 
power-sensitive circuits. In this mode, power is swept at a single frequency, from a start 
power value to a stop power value, selected using the 
 and 
 keys and the 
entry block. This feature is convenient for such measurements as gain compression or AGC 
(automatic gain control) slope. To set the frequency of the power sweep, use 
 in 
the stimulus menu.
The span of the swept power is limited to being equal to or within one of the eight 
pre-defined power ranges. The attenuator will not switch to a different power range while 
in the power sweep mode. Therefore, when performing a power sweep, power range 
selection will automatically switch to the manual mode.
In power sweep, the entered sweep time may be automatically changed if it is less than the 
minimum required for the current configuration (number of points, IF bandwidth, 
averaging, etc.).
CW Time Sweep (Seconds) 
The 
 softkey turns on a sweep mode similar to an oscilloscope. The analyzer is 
set to a single frequency, and the data is displayed versus time. The frequency of the CW 
time sweep is set with 
 in the stimulus menu. In this sweep mode, the data is 
continuously sampled at precise, uniform time intervals determined by the sweep time and 
the number of points minus 1. The entered sweep time may be automatically changed if it 
is less than the minimum required for the current instrument configuration.
In time domain using Option 010, the CW time mode data is translated to frequency 
domain, and the x-axis becomes frequency. In this mode, the instrument can be used as a 
spectrum analyzer to measure signal purity, or for low frequency (<1 kHz) analysis of 
amplitude or pulse modulation signals.
Selecting Sweep Modes 
In addition to the previous sweep types, there are also two different sweep modes. These 
can be accessed through the correction menu by pressing 
 
 
 or 
SEGMENT IF BW 
LIST IF BW ON off 
POWER SWEEP 
Start
Stop
CW FREQ 
CW TIME 
CW FREQ 
Cal
MORE 
ALTERNATE A and B 
CHOP A and B