Operating Concepts
Limit Line Operation
If limit lines are on, they are plotted with the data on a plot. If limit testing is on, the 
PASS
 
or 
FAIL
 message is plotted, and the failing portions of the trace that are a different color on 
the display are also a different color on the plot. If limits are specified, they are saved in 
memory with an instrument state.
Edit Limits Menu 
This menu allows you to specify limits for limit lines or limit testing, and presents a table 
of limit values on the display. Limits are defined in segments. Each segment is a portion of 
the stimulus span. Up to 22 limit segments can be specified for each channel. The limit 
segments do not have to be entered in any particular order: the analyzer automatically 
sorts them and lists them on the display in increasing order of start stimulus value.
For each segment, the table lists the segment number, the starting stimulus value, upper 
limit, lower limit, and limit type. The ending stimulus value is the start value of the next 
segment, or a segment can be terminated with a single point segment. You can enter limit 
values as upper and lower limits or delta limits and middle value. As new limit segments 
are defined, the tabular listing is updated. If limit lines are switched on, they are shown on 
the display.
If no limits have been defined, the table of limit values shows the notation 
EMPTY
. Limit 
segments are added to the table using the 
 softkey or edited with the 
 softkey, 
as previously described. The last segment on the list is followed by the notation 
END
.
Edit Segment Menu 
This menu sets the values of the individual limit segments. The segment to be modified, or 
a default segment, is selected in the edit limits menu. The stimulus value can be set with 
the controls in the entry block or with a marker (the marker is activated automatically 
when this menu is presented). The limit values can be defined as upper and lower limits, or 
delta limits and middle value. Both an upper limit and a lower limit (or delta limits) must 
be defined: if only one limit is required for a particular measurement, force the other out of 
range (for example +500 dB or 
500 dB).
As new values are entered, the tabular listing of limit values is updated.
Segments do not have to be listed in any particular order: the analyzer sorts them 
automatically in increasing order of start stimulus value when the 
 key in the edit 
limits menu is pressed. However, the easiest way to enter a set of limits is to start with the 
lowest stimulus value and define the segments from left to right of the display, with limit 
lines turned on as a visual check.
Phase limit values can be specified between +500
°
 and 
500
°
. Limit values above +180
°
 
and below 
180
°
 are mapped into the range of 
180
°
 to +180
°
 to correspond with the range 
of phase data values.
Offset Limits Menu 
This menu allows the complete limit set to be offset in either stimulus value or amplitude 
value. This is useful for changing the limits to correspond with a change in the test setup, 
or for device specifications that differ in stimulus or amplitude. It can also be used to move 
the limit lines away from the data trace temporarily for visual examination of trace detail.