Operating Concepts
Knowing the Instrument Modes
Knowing the Instrument Modes 
There are three major instrument modes of the analyzer:
• network analyzer mode
• tuned receiver mode
• frequency offset operation (Option 089)
Network Analyzer Mode 
This is the standard mode of operation for the analyzer, and is active after you press 
 or switch on the AC power. This mode uses the analyzer's internal source.
Pressing 
 
 
 returns the analyzer 
to the "normal" network analyzer operating mode.
Tuned Receiver Mode
In the tuned receiver mode, the analyzer's receiver operates independently of any signal 
source. This mode is not phase-locked and functions in all sweep types. The analyzer tunes 
the receiver to a synthesized CW input signal at a precisely specified frequency. All phase 
lock routines are bypassed, increasing sweep speed significantly. The external source must 
be synthesized, and must drive the analyzer's external frequency reference. The analyzer's 
internal source frequency is not accurate, and the internal source should not be used in the 
tuned receiver mode.
Using the analyzer's tuned receiver mode is useful for automated test applications where 
an external synthesized source is available and applications where speed is important. 
Although the tuned receiver mode can function in all sweep types, it is typically used in 
CW applications.
Typical test setup 
1. Activate the tuned receiver mode by pressing 
 
 
.
2. To perform a CW measurement using the tuned receiver mode, connect the equipment 
as shown in 
.
Preset
System
INSTRUMENT MODE  NETWORK ANALYZER 
System
INSTRUMENT MODE 
TUNED RECEIVER