Agilent Technologies E1441A Manuale Utente

Pagina di 206
Service Procedures  
167
Appendix D
Quick Performance
Check
The quick performance check is a combination of internal self-test and an 
abbreviated performance test (specified by the letter 
Q
 in the performance 
verification tests). This test provides a simple method to achieve high 
confidence in the function generator's ability to functionally operate and meet 
specifications. These tests represent the absolute minimum set of 
performance checks recommended following any service activity. Auditing 
the function generator's performance for the quick check points (designated 
by a 
Q
) verifies performance for “normal” accuracy drift mechanisms. This 
test does not check for abnormal component failures.
To perform the quick performance check, do the following:
Set the function generator to reset state (*RST).
Perform a complete self-test (*TST?).
Perform only the performance verification tests indicated with the 
letter 
Q
.
If the function generator fails the quick performance check, adjustment or 
repair is required.
Performance
Verification Tests
The performance verification tests are recommended as acceptance tests 
when you first receive the function generator. The acceptance test results 
should be compared against the 1 year test limits. After acceptance, you 
should repeat the performance verification tests at every calibration interval.
If the function generator fails performance verification, adjustment or repair 
is required.
Frequency Verification
This test verifies the frequency accuracy of the two sources in the function 
generator. All output frequencies are derived from a single generated 
frequency, and only one frequency point is checked. The second test verifies 
the burst rate frequency.
Set the function generator for each output indicated in the table below. Use 
a frequency meter to measure the output frequency. Compare the measured 
results to the test limits shown in the table. This is a 50 
Ω  output termination 
test.
Table D-2. Frequency Verification
Agilent E1441A
Measurement
Function
 Out
Term
a
Ampl
Freq
Burst
 Rate
Burst
Count
Nominal
Error 
Q
Sine wave
50 
Ω
3.5 Vrms
1.00 kHz
1.00 kHz
± 
0.01 Hz
Q
Square wave
50 
Ω
 
3.5 Vrms
1.00 kHz
500 Hz
1 CYC
500 Hz
± 
5 Hz
a.The E1441A has a fixed output impedance of 50 
Ω
 on the "Output" terminal.  Use the 
OUTPut:LOAD 50 | INFinity | MIN | MAX command to set the output termination. HIGH Z 
assumes no load on output. 50 
Ω
 assumes a 50 
Ω
 
±
 0.1 
Ω
 load on output.