Agilent Technologies E1441A Manuale Utente

Pagina di 206
Service Procedures  
169
Appendix D
AC Amplitude Verification
This procedure is used to check the output amplitude calibration of the 
function generator. Verification checks are performed to check the accuracy 
of the pre-attenuator and post attenuator. Make sure you have read “Test 
Considerations” on page 165.
Set the function generator for each output indicated in the table on the next 
page. Use a DMM to measure the ACrms output voltage of the function 
generator. Compare the measured results to the test limits shown in the table. 
This is a HIGH Z output termination test.
Table D-5. AC Amplitude Verification (High Z)
Agilent E1441A
Measurement
Function
 Out
Term
a
Ampl
Freq
Nominal
Error 
Q Sine wave
HIGH Z
7.0 Vrms
1.00 kHz
7.0 Vrms
± 
0.070 Vrms
Sine wave
HIGH Z
5.7 Vrms
1.00 kHz
5.7 Vrms
± 
0.057 Vrms
Sine wave
HIGH Z
5.5 Vrms
1.00 kHz
5.5 Vrms
± 
0.055 Vrms
Sine wave
HIGH Z
4.4 Vrms
1.00 kHz
4.4 Vrms
± 
0.044 Vrms
Sine wave
HIGH Z
3.5 Vrms
1.00 kHz
3.5 Vrms
± 
0.035 Vrms
Sine wave
HIGH Z
2.8 Vrms
1.00 kHz
2.8 Vrms
± 
0.028 Vrms
Sine wave
HIGH Z
2.2 Vrms
1.00 kHz
2.2 Vrms
± 
0.022 Vrms
Sine wave
HIGH Z
1.7 Vrms
1.00 kHz
1.7 Vrms
± 
0.017 Vrms
Sine wave
HIGH Z
1.4 Vrms
1.00 kHz
1.4Vrms
± 
0.014 Vrms
Sine wave
HIGH Z
1.1 Vrms
1.00 kHz
1.1 Vrms
± 
0.011 Vrms
Q Sine wave
HIGH Z
0.88 Vrms
1.00 kHz
0.88 Vrms
±
 0.0088 Vrms
Sine wave
HIGH Z
0.70 Vrms
1.00 kHz
0.70 Vrms
± 
0.0070 Vrms
Sine wave
HIGH Z
0.55 Vrms
1.00 kHz
0.55 Vrms
± 
0.0055 Vrms
Sine wave
HIGH Z
0.44 Vrms
1.00 kHz
0.44 Vrms
± 
0.0044 Vrms
Sine wave
HIGH Z
0.35 Vrms
1.00 kHz
0.35 Vrms
± 
0.0035 Vrms
Sine wave
HIGH Z
0.28 Vrms
1.00 kHz
0.28 Vrms
± 
0.0028 Vrms
Sine wave
HIGH Z
0.22 Vrms
1.00 kHz
0.22 Vrms
±
 0.0022 Vrms
Sine wave
HIGH Z
0.17 Vrms
1.00 kHz
0.17 Vrms
±
 0.0017 Vrms
Sine wave
HIGH Z
0.14 Vrms
1.00 kHz
0.14 Vrms
± 
0.0014 Vrms
Sine wave
HIGH Z
0.11 Vrms
1.00 kHz
0.11 Vrms
±
 0.0011 Vrms
Q Sine wave
HIGH Z
0.088 Vrms
1.00 kHz
0.088Vrms
± 
0.00088 Vrms
Sine wave
HIGH Z
0.070 Vrms
1.00 kHz
0.070 Vrms
± 
0.00070 Vrms
Sine wave
HIGH Z
0.055 Vrms
1.00 kHz
0.055 Vrms
± 
0.00055 Vrms
Sine wave
HIGH Z
0.044 Vrms
1.00 kHz
0.044 Vrms
± 
0.00044 Vrms
Q Sine wave
HIGH Z
0.036 Vrms
1.00 kHz
0.036 Vrms
± 
0.00036 Vrms
a.The E1441A has a fixed output impedance of 50 
Ω
 on the "Output" terminal.  Use the 
OUTPut:LOAD 50 | INFinity | MIN | MAX command to set the output termination. 
HIGH Z assumes no load on output. 50 
Ω
 assumes a 50 
Ω
 
±
 0.1 
Ω
 load on output.