Agilent Technologies E1441A Manuale Utente
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
168
Service Procedures
Appendix D
Function Gain and Linearity Verification
This test verifies the output amplitude accuracy specification for
sine wave, triangle wave, ramp, and square wave outputs.
sine wave, triangle wave, ramp, and square wave outputs.
Set the function generator for each output indicated in the table below. Use
a DMM to measure the function generator ACrms output voltage. Compare
the measured results to the test limits shown in the table. This is a HIGH Z
output termination test.
a DMM to measure the function generator ACrms output voltage. Compare
the measured results to the test limits shown in the table. This is a HIGH Z
output termination test.
DC Function Offset Verification
This test verifies the DC offset and DC output specifications.
Set the function generator for each output indicated in the table below. Use
a DMM to measure the function generator DCV output. Compare the
measured results to the test limits shown in the table. This is a HIGH Z output
termination test.
a DMM to measure the function generator DCV output. Compare the
measured results to the test limits shown in the table. This is a HIGH Z output
termination test.
Table D-3. Function Gain and Linearity Verification
Agilent E1441A
Measurement
Function
Out
Term
a
Ampl
Freq
Nominal
Error
Q
Sine wave
HIGH Z
7.0 Vrms
1.0 kHz
7.0 Vrms
±
0.07 Vrms
Sine wave
HIGH Z
5.7 Vrms
1.0 kHz
5.7 Vrms
±
0.057 Vrms
Triangle wave
HIGH Z
5.7 Vrms
100 Hz
5.7 Vrms
±
0.057 Vrms
Ramp wave
HIGH Z
5.7 Vrms
100 Hz
5.7 Vrms
±
0.057 Vrms
Q
Square wave
HIGH Z
10.0 Vrms
100 Hz
10.0 Vrms
±
0.1 Vrms
Square wave
HIGH Z
8.0 Vrms
100 Hz
8.0 Vrms
±
0.08 Vrms
a.The E1441A has a fixed output impedance of 50
Ω
on the "Output" terminal. Use
the OUTPut:LOAD 50 | INFinity | MIN | MAX command to set the output
termination. HIGH Z assumes no load on output. 50
termination. HIGH Z assumes no load on output. 50
Ω
assumes a 50
Ω
±
0.1
Ω
load on output.
Table D-4. DC Function Offset Verification
Agilent E1441A
Measurement
Function
Out
Term
a
a.The E1441A has a fixed output impedance of 50
Ω
on the "Output"
terminal. Use the OUTPut:LOAD 50 | INFinity | MIN | MAX command to
set the output termination. HIGH Z assumes no load on output. 50
set the output termination. HIGH Z assumes no load on output. 50
Ω
assumes a 50
Ω
±
0.1
Ω
load on output.
Ampl
Nominal
Error
Q
DC Volts
HIGH Z
10.0 VDC
10.0 VDC
±
0.20 VDC
DC Volts
HIGH Z
-10.0 VDC
-10.0 VDC
±
0.20 VDC