8.2. 
The Precision Weight Test  
 
The precision weight test consists of performing the weight test in Section 8.1 multiple 
times to determine the instruments repeatability.  This is a more critical verification of the 
instrument’s performance. 
Mean, Standard Deviation, and Relative Standard Deviation are calculated using the 
following formulas. 
Mean = total of results divided by number of results. 
Standard Deviation = 
1
-
n
x
-
x
2
)
(
 
Relative Standard Deviation expressed as percentage = Standard Deviation 
divided by Mean, multiplied by 100 
 
  Perform the steps in the weight 
test a total of five (5) times. 
  After all five tests are complete, 
press [MENU] and select the 
STORED DATA MENU. 
  PRESS [ENT] to select tests. 
  Arrow up or down to select the 
five weight tests and press 
[ANALYZE]
  The mean of the results should 
be 37.500%, +/-.020 for a test 
using the five and three gram 
weights. 
  If the results obtained from these 
tests are outside the specified 
range, call AZI Customer Service 
at 800-528-7411 or 602-470-1414 
for assistance.