8.3. 
Sodium Tartrate Test  
 
 
The sodium tartrate test verifies the heater and force balance operation.  Sodium tartrate 
dihydrate is a chemical with a known moisture concentration that releases moisture at 
temperatures greater than 120º C. 
 
Because of its consistent moisture concentration, it is used as a standard in moisture 
measurement using oven or titration methods.  To obtain consistently reliable results, use 
a high quality or reagent grade (ACS certified) granular sodium tartrate.  Sift the sodium 
tartrate through a #20 sieve to obtain a more uniform size that will achieve better results. 
Lumps in sodium tartrate usually indicate moisture absorption that will yield inconsistent 
results.  Discard the larger chunks, or pestle them through the sieve and return them to 
the bottle. 
 
Sodium tartrate is available from AZI as P/N 800-0037.  Call AZI part sales at 
800-528-7411 or 602-470-1414 for information or to order. 
 
The sodium tartrate test uses approximately 8 grams of sample and is run at 150º
 
C.  
Results should be within a range of ±0.1 of the assayed value on the label of the chemical. 
 Sodium tartrate assayed moistures typically range from 15.61 - 15.71. 
  Open the lid; place a clean, dry, empty pan the pan 
support 
  Close the lid 
  From main test screen, press [MEMORY]  
  Highlight "(Factory) Tartrate" and press [START], then 
press [ABORT]   (This will set the instrument idle 
temperature to 80 degrees.) 
  Allow the instrument to reach 80°C (Approx. 3-5 minutes) 
  When the instrument has reach 80°C, press [START] 
  Allow instrument to tare empty pan 
  When instructed, open the lid 
  Add sample; use a spoon to evenly spread 7.9 to 8.1 
grams of Tartrate on the surface of the pan (8 grams is 
best) 
  Close the lid and allow instrument to start the test 
  Allow the test to complete 
  If this is the first test of the day, discard the results of this test for stability reasons.  
  Results should be within a range of ±0.1 of the assayed value on the bottle label.