Section 6: Meter operation
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
6.5.5
GC data validity
The GC-read gas property data validity is indicated by the 
AreGasPropertiesInvalidGC data 
point. The data is considered invalid if any of the conditions listed in 
 is true. These 
conditions are indicated by the Daniel MeterLink 
Monitor
 page 
Field I/O
 indicator.
6.5.6
GC alarm handling
The following data points are included in the meter’s alarm log (
Are-
GasPropertiesInvalidGCIsGCAlarmPresentIsGCCommErrGCCommStatus, and 
IsGCDataErr.
19
GC response message (RTU protocol) incomplete.
20
GC gas property data spans more than one update.
Table 6-17  GC-read gas property invalid conditions
Indication
Data Point
Condition
IsGCAlarmPresent
Invalid or expired GC feature key (
).
Option Board Port C is not available (Option Board probably disconnected).
Option Board Port C improperly configured (such as if configured without using the Daniel 
MeterLink Field Setup Wizard).
Unrecognized gas component ID.
Specified gas stream (
GCStreamNumber
) not found within the specified period (
GCDe-
siredStreamTimeout
).
GC reported alarm (GC Alarm1 bits 14 and/or 15 set, GC Alarm2 bits 0, 1, 2, and/or 3 set).
IsGCWarningPresent
Not presently used - reserved for future use.
IsGCCommErr
Refer to communication error conditions listed in 
.
IsGCDataErr
The total unnormalized gas composition mole percentage is not within [85%, 115%].
An individual gas composition mole percentage is not within [0%, 100%].
The specific gravity is not within [0.2, 0.8].
The heating value is greater than 50 kJ/dm
3
.
Table 6-16  GC Communication status list
GCComm status value
Error description