Whenever the analyzer starts or finishes a calibration, it issues a status report to the RS-232 
interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise, 
they will be discarded. The format of these messages is: 
C DDD:HH:MM IIII CALIBRATION STATUS MESSAGE 
An example of an actual sequence of calibration status messages is: 
C DDD:HH:MM IIII START MULTI-POINT CALIBRATION 
C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION 
There are several methods of both checking the calibration and calibrating the M200AH, these 
are discussed in Section 7. C LIST lists the calibration commands available. The C command 
executes a calibration command, which may be one of the following: 
Table 5-18:  RS-232 Calibration Commands 
Command 
Description 
C LIST 
List calibration commands 
C ZERO 
Enter zero calibration mode 
C SPAN 
Enter span calibration mode 
C COMPUTE ZERO 
Do a zero calibration 
C COMPUTE SPAN 
Do a span calibration 
C EXITZ 
Exit zero calibration 
C EXITS 
Exit span calibration 
C EXIT 
Exit zero or span calibration 
 
5.7.6.   DIAGNOSTIC Commands and Messages 
When Diagnostic mode is entered from the RS-232 port, the diagnostic mode issues additional 
status messages to indicate which diagnostic test is currently selected. Examples of Diagnostic 
mode messages are: 
D DDD:HH:MM IIII ZERO VALVE=ON 
D DDD:HH:MM IIII ENTER DIAGNOSTIC MODE 
D DDD:HH:MM IIII EXIT DIAGNOSTIC MODE