Teledyne 200AH 用户手册

下载
页码 196
Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
5-39 
The following is a summary of the Diagnostic commands. 
Table 5-19:  RS-232 Diagnostic Command Summary 
Command 
Description 
D LIST 
D NAME=VALUE 
Examines or sets I/O signal. Must issue D ENTER SIG command 
before using this command. 
D LIST NAMES 
Prin is command.ts names of all diagnostic tests. 
D ENTER SIG 
D ENTER OT 
D ENTER ET 
Executes SIGNAL I/O diagnostic test. 
Executes Optic Test diagnostic test. 
Executes Elect Test diagnostic test. 
Use D EXIT to leave these diagnostic modes. 
D EXIT 
Must use this command to exit SIG, ET or OT Diagnostic modes. 
D SYS_RESET 
Resets analyzer software (same as power on). 
D RAM_RESET 
Resets analyzer software and erases RAM. Erases NO, NO
X
, NO
2
 conc 
values, erases all DAS data. Keeps setup variables and calibration. 
(Same as installing new software version.) 
D EE_RESET 
Resets analyzer software and erases RAM and EEPROM. Returns all 
setup variables to factory defaults, resets calibration values. 
 
5.7.7.   DAS Commands and Messages 
The M200AH contains a flexible and powerful built in data acquisition system (DAS) that 
enables the analyzer to store concentration data as well as diagnostic parameters in its battery 
backed memory. This information can be printed out through the RS-232 port. The diagnostic 
data can be used for performing “Predictive Diagnostics” and trending to determine when 
maintenance and servicing will be required. 
To print out the properties of all of the data channels enter: 
D PRINT 
To print the properties of just a single data channel enter: 
D PRINT "name” 
For example to print the properties of the CONC data channel enter: 
D PRINT “CONC”