Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
6-3 
6.2.1.   Autocal - Setup Zero/Span Valves 
The Zero/Span valves can be set up to operate automatically. Table 6-2 in this section describes 
Autocal setup. 
Table 6-2:  Setup Automatic Zero/Span Checking or Calibration 
Step 
Action 
Comment 
1. 
Press SETUP-ACAL 
This button sequence will cause the AUTOCAL menu to be displayed. 
2. 
Press PREV or 
NEXT 
This selects a SEQUENCE. Up to 3 Sequences can be defined. A 
Sequence is a definition of zero/span calibrations along with timing 
information. 
3. 
Select MODE, then 
PREV or NEXT 
MODE allows you to select the order of zero, low-span or high-span 
that will be performed, MODES are listed in Table 6-3. Use PREV - 
NEXT to scroll the list of choices, then press ENTR to select. To 
disable a sequence, select DISABLE from the menu.  
4. 
Press SET 
This menu sets up several timers and selects calibration or calibration 
check. 
1.  TIMER ENABLE – If set to ON, the defined sequence will 
execute. If set to OFF, the sequence will remain in memory, but 
will not execute. 
2.  START DATE - calendar date to start sequence dd-mmm-yy, day-
month-year format.  
3.  START TIME - time of day sequence will start. hh-mm, hours-
minutes in 24 hr format.  
4.  DELTA DAYS and DELTA HOURS:MIN is the time between 
each activation of the sequence. By setting the Delta Hours 
variable to be other than an exact number of days, the sequence 
can move ahead or be delayed each time it is run. For example, by 
setting DELTA DAYS = 1 and DELTA HOURS = 1:00, the 
sequence will execute 1 hour later each day. To set it to move back 
½ hour each day, DELTA DAYS = 0, DELTA HOURS = 23:30. 
5.  DURATION -Amount of time for each zero/span point. 
6.  CALIBRATE – If set to ON, will calibrate(adjust slope and 
offset). If OFF will check, but not change calibration. 
7.  7. RANGE-TO-CAL Selects the range to calibrate LO or HI. 
Applicable to AutoRange and RemoteRange modes. 
5. 
Press EXIT 
The M200AH now exits the AUTOCAL SETUP menu. For each 
enabled sequence, the M200AH will now AUTOCAL at the defined 
times.