6.4. 
  Isolated 4-20 mA Current Loop Option 
The M200AH comes equipped with 4-20mA current loop. The electrical ground of this output is 
part of the instrument ground. In certain applications the instrument ground must be isolated 
from the external ground. For applications requiring electrical isolation, we offer optional 
isolated current loop outputs. With the exception of electrical isolation, the setup and usage is 
identical to the standard equipment current loop output. The calibration procedure is located in 
the Troubleshooting Section 9. 
6.5. 
  Molybdenum Converter Option 
The Molybdenum Converter is used in applications where the NO
2
 concentration will not be 
expected to exceed 5 ppm. In these applications it exhibits very high conversion efficiencies, 
thus enabling the accurate measurement of low concentrations of NO
2
. Operating the Converter 
where the concentrations exceed 5 ppm will cause lower efficiencies and will shorten the life of 
the converter. 
6.6. 
  External Desiccant Canister Option 
This Option is located between the PermaPure Dryer and the ozone flowmeter. See Figure 2-2 
for location and Figure 8-5 for pneumatic schematic. It is mounted external to the analyzer. It 
should be used at extremely humid sites, or where the -70°C dewpoint air produced helps in 
reducing reaction cell contamination and window clouding. 
6.7. 
  Alternate Bypass Flow Orifice Option 
The standard bypass flow is 250 cc/min. For those applications requiring a higher 
sample flow, this option provides a 500 cc/min bypass flow. The overall sample flow is 
550 cc/min with this option installed. The orifice is located in the vacuum manifold at 
the rear of the instrument see  
Figure 2-6. 
6.8. 
  M501 External Converter Option 
The M200AH can be ordered with a Model 501 Thermal Converter. Its conversion performance 
is very similar to the Mini HiCon. The advantage of this converter is that it is less prone to 
plugging because it is constructed of larger diameter tubing, also the conversion efficiency is 
very high because the thermal converter is followed by a molybdenum converter. Since the 
converter is external to the M200AH the response of the instrument is considerably slower if 
operated in the NO/NO
X
 switching mode. This is due to the larger internal volume that must be 
purged during the NO/NO
X
 measurement cycle. This effect is less noticeable if the M200AH is 
being operated in the NO
X
 only mode.