Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
6-4 
Table 6-3:  Action of MODE Field in AutoCal 
MODE 
CALIBRATE = ON 
CALIBRATE = OFF 
ZERO 
Zero calibrates LO or HI (depending 
on RANGE-TO-CAL), SINGLE or 
IND ranges. 
Zero checks LO or HI (depending on 
RANGE-TO-CAL), SINGLE or IND 
range. 
ZERO-LO 
Zero and low span calibrates LO, 
SINGLE or IND ranges. 
Zero and low span checks LO, SINGLE 
or IND ranges. 
ZERO-HI 
Zero cals & span cals HI, SINGLE or 
IND range with HI span gas. 
Zero checks & span checks on HI, 
SINGLE or IND range with HI span 
gas. 
ZERO-LO-HI 
Zero cals, span cals LO and HI range. 
LO span gas for LO range, HI span 
gas for HI range. 
Zero cals & span cals in SINGLE or 
IND range using HI span gas. LO 
span is cal check only. 
Zero checks, span checks both ranges. 
LO span gas for LO range, HI span gas 
for HI range. Does 3 point check in 
SINGLE or IND range modes. 
LO-HI 
Span cals LO range with LO span gas, 
Span cals HI range with HI span gas 
in AutoRange mode. Span cals 
SINGLE or IND range with HI span 
gas. LO span is cal check only. 
Span checks LO range with LO span 
gas, Span checks HI range with HI span 
gas. Does 2 point check in SINGLE or 
IND range mode. 
LO 
Span cals LO range with LO span gas. 
Span cals SINGLE or IND range with 
LO span gas. 
Span checks LO range with LO span 
gas. Span checks SINGLE or IND range 
with LO span gas. 
HI 
Span cals on HI range with HI span 
gas. Span cals SINGLE or IND range 
with HI span gas. 
Span checks on HI range with HI span 
gas. Span checks SINGLE or IND range 
with HI span gas. 
 
6.3. 
  Oxygen Sensor Option 
The Model 200AH can be equipped with an optional paramagnetic oxygen sensor that is capable 
of measuring full scale ranges from 5% to 100%.  The option consists of the paramagnetic sensor 
module, an interface electronics board, a flow control orifice that generates a flow of about 
80 cc/min through the sensor, and a temperature controlled oven that maintains the sensor and 
flow control orifice at 50
C.  
Note that an Oxygen Sensor cannot be included in an instrument that also includes the Zero/Span 
Valves option.