8. CALIBRATION 
‰ Calibration of the scale is required when it is initially installed, if it is moved often, or 
it is moved a substantial distance. Calibration is also necessary in regular scale 
maintenance due to normal mechanical wear-and-tear, changes in seasonal 
temperature, humidity, air pressure, etc. 
The scale is equipped with gravity compensation, which allows it to be calibrated in 
one location and then adjusted to match the gravity acceleration at another location 
where it will be used. But don’t worry about this, as far as you calibrate the scale 
using a calibration weight and use it at same place. 
The scale must perform "warm up" for at least 30 minutes before starting 
calibration.  
8-1. Calibration Procedure Using a Weight 
The scale should be powered on at least one-half hour to warm it up before starting 
the calibration procedure. 
1. Remove the calibration switch cover, and press the calibration (CAL) switch. 
  The scale shows “Cal” in the count display. 
‰ 
Press the  ON/OFF  key to exit 
without calibrating the scale. 
‰ Press and hold the  PRINT  key and 
press the  ON/OFF  key, then you can 
also enter calibration mode. 
2. Press the  ZERO  key to enter into zero & 
span calibration mode. 
‰ The display flashes the required calibration 
weight value. 
Calibration (CAL)
switch
Bottom view
Calibration (CAL)
switch cover 