A&D HC-6Ki Benutzerhandbuch

Seite von 60
29
8-2. Gravity Compensation 
When the scale is first used or has been moved to different place, it should be 
calibrated using a calibration weight. 
But if the calibration weight cannot be prepared, the gravity acceleration correction 
will compensate the scale. Change the gravity acceleration value of the scale to the 
value of the area where it will be used. Refer to the gravity acceleration map 
appended to the end of this manual.   
1. In the Step 1 above, press  TARE   key 
The display flashes the gravity acceleration 
value stored in the scale. 
2. Use the  0
Æ
9  10-key pad to 
display the desired gravity 
acceleration value. 
(Example of the value 9.800 m/s
2
.)
3. Press the  ENTER   key. 
The scale will store the new value. 
If necessary to calibrate the scale 
using a weight, go to Step 2 of the 
previous section. 
4. Press the  ON/OFF  key to turn the scale off and re-attach the calibration switch 
cover. (End of the calibration procedure.) 