A&D HC-6Ki Benutzerhandbuch

Seite von 60
28
If you know the exact weight value, or 
if you wish to use a different weight,   
use the  0
Æ 9  and   10-key pad to 
display the desired calibration weight. 
(For example: Using 5 kg calibration weight 
that actually weighs 5.001 kg. Do not forget 
to enter a decimal point in this case.) 
3. Press the  ENTER   key. 
The calibration weight stops blinking. 
4.  Making sure that there is nothing on, or touching 
the weighing pan, press the  ENTER   key. 
When zero calibration is completed, the display 
will show “Cal f”. 
If you don’t need span calibration, press the 
ON/OFF  key to exit from the calibration 
procedure.
5. Place the calibration weight on the weighing pan 
and press the    ENTER    key. 
When span calibration is completed, the display 
returns to step 1 showing the weight value for the 
calibration weight. 
Remove the calibration weight. 
If the calibration weight is not what it should be, 
an error will be displayed. Check if the weight is 
correct and try again. 
6. Press the  ON/OFF  key to turn the scale off 
and re-attach the calibration switch cover. 
(End of the calibration procedure.) 
If the scale will be moved to another place, set the gravity
acceleration value before calibration. The value must be of the
area where the calibration is to be done. 
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
Cal
0
0
8
8
0.0
5.001
0
00
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
Cal
0
0
8
8
0.0
5.000
0
00
ENTER
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
Cal
0
f
8
8
0.0
5.000
0
00
Nothing on the 
weighing pan!
ENTER
888
Cal
02
COUNT
WEIGHT UNIT WEIGHT
88-
5.000
8
8
11Hi00
Place
Calibration 
i ht
ENTER