4.1.7 Page 7: Monitor Settings
Picture 4-8: monitor settings
Name
Term
Range
Default Remarks
NFiles
Reader data onto disk
1 ... 999
1
AntDisplay
Antenna to monitor
0 .. 2
1
0= both Antennas, 1=Ant1, 2=Ant2
TrigAnt
Antenna to trigger
1 .. 3
3
1=Ant.1, 2=Ant2, 3=both Antennas
Table 4-8: monitor settings
1
Nfiles
Defines the number of „pictures“ of a current identification situation which will be stored onto disk when
using the function „save reader data onto disk“ (see chapter Remote Diagnostics on page 64).
using the function „save reader data onto disk“ (see chapter Remote Diagnostics on page 64).
AntDisplay
Select the antenna (A) which you want to look at with the service PC:
•
AntDisplay=0
the data of the antennas 1 and 2 are displayed alternately
•
AntDisplay=1
only the data of antenna 1 is displayed
•
AntDisplay=2
only the data of antenna 2 is displayed
TrigAnt
Select the antenna (A) which you want to trigger with the help of key „t“ on your service PC.
•
TrigAnt=1
only antenna 1 is selected to be triggered
•
TrigAnt=2
only antenna 2 is selected to be triggered
•
TrigAnt=3
antennas 1 and 2 are selected to be triggered
1 These two parameters are not saved in the reader (R
)
. They are only used in the service PC..