Boston Scientific Corporation CRM70514 Benutzerhandbuch

Seite von 86
37
Q-TECH™ PROGRAMMER:  OPERATiOn
9. 
Once the Reference S-ECG acquisition process begins, a status screen appears. The 
process may take up to one minute, during which the patient should remain still. 
During this process, a template of the patient’s baseline QRS complex is stored in the 
SQ-RX device. Select Cancel at any time to end Reference S-ECG acquisition. When 
acquisition is complete, select the Continue button.
Programming Therapy Parameters
Once Automatic Setup has been completed, the SQ-RX device therapy parameters may be selected.
Note:   The IDE Study demonstrated a significant reduction in inappropriate 
therapy with the activation of the Conditional Shock Zone prior to 
hospital discharge  (see SQ‐RX Pulse Generator User’s Manual, Section: 
S‐ICD System Clinical Investigation).
To set the therapy parameters:
1. 
Select the Main Menu icon on the Navigation Bar.
2. 
Select the Implant button.
3. 
Select the Device Settings icon on the Navigation Bar to display the Device Settings 
Figure 29:  Device Settings screen