37
Q-TECH™ PROGRAMMER: OPERATiOn
9.
Once the Reference S-ECG acquisition process begins, a status screen appears. The
process may take up to one minute, during which the patient should remain still.
During this process, a template of the patient’s baseline QRS complex is stored in the
SQ-RX device. Select Cancel at any time to end Reference S-ECG acquisition. When
acquisition is complete, select the Continue button.
Programming Therapy Parameters
Once Automatic Setup has been completed, the SQ-RX device therapy parameters may be selected.
Note: The IDE Study demonstrated a significant reduction in inappropriate
therapy with the activation of the Conditional Shock Zone prior to
hospital discharge (see SQ‐RX Pulse Generator User’s Manual, Section:
S‐ICD System Clinical Investigation).
To set the therapy parameters:
1.
Select the Main Menu icon on the Navigation Bar.
2.
Select the Implant button.
3.
Select the Device Settings icon on the Navigation Bar to display the Device Settings
screen (Figure 29).
Figure 29: Device Settings screen